Quantitative 3w-scanning thermal microscopy : Modelling the AC/DC coupling and the sample heat conduction

Show full item record

Files in this item

PDF Q3O210.pdf 190.4Kb

Pour citer ce document :
URI: http://hdl.handle.net/2042/6572
Title: Quantitative 3w-scanning thermal microscopy : Modelling the AC/DC coupling and the sample heat conduction
Author: Chapuis, P.-O.; Kumar Saha, S.; Volz, S.
Abstract: A way to increase the Scanning Thermal Microscope (SThM) sensitivity in the harmonic 3w mode is to heat the probe with an AC current sufficiently high to generate a coupling between the AC and the DC signals. We detail in this paper how to properly take into account this coupling with a Wollaston-probe SThM. We also show how to link correctly the thermal conductivity to the thermal conductance measured by the SThM.
Subject: Scanning thermal microscopy (SThM), Thermal conductivity
Publisher: TIMA Editions , Grenoble, France